Command reference
R&S
®
ZNA
1173
User Manual 1178.6462.02 ─ 20
Example:
*RST; :CALC2:PAR:SDEF 'TRC2','S11'
DISPlay:WINDow:TRACe2:FEED 'Trc2'
Reset the analyzer to create the default channel no. 1 and
default trace. Create a second trace and display the trace. Both
traces are measured continuously.
INIT1:CONT OFF
Activate single sweep mode for the first channel. The measure-
ment in channel no. 1 is stopped after the current sweep. Chan-
nel no. 2 is still measured continuously.
INIT1
Activate a new (single) sweep in channel no. 1. Channel no. 2 is
still measured continuously.
INIT:CONT:ALL OFF
Activate single sweep mode for all channels. The measurement
in channel no. 2 is also stopped after the current sweep.
INIT:ALL
Re-start a single sweep in both channels.
Example:
Alternative settings using the R&S ZVAB compatibility mode:
*RST; :SYSTem:LANGuage 'ZVABT'
*RST; :INIT:CONT OFF
Activate single sweep mode for all channels (including channel
no. 2 created later).
INIT:SCOP SING
State that a single sweep will be performed in the active chan-
nel.
CALC2:PAR:SDEF 'TRC2','S11'; :INIT2
Create channel no. 2 with a new trace and start a single sweep
in channel no. 2. Start a single sweep in the second channel.
Manual operation:
See
INITiate<Ch>:CTIMing
<Boolean>
Activates/deactivates the controlled timing mode.
Can only be executed if the current setup contains a single channel.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
Options:
R&S
ZNA-K28
Manual operation:
See
SCPI command reference