GUI reference
R&S
®
ZNA
544
User Manual 1178.6462.02 ─ 20
In the converter configuration above, P1 and P2 drive the RF IN ports of converter
1 and 2, respectively, and the Converter LO port drives the LO IN ports of both con-
verters.
Since firmware version 2.40, the "Leveling Table" buttons give access to the con-
verter leveling functionality that previously required a separate tool.
Select a "Ref. Receiver" or "Meas. Receiver", "Leveling Table", or "Source Flatness"
symbol to start the related calibration. See
"Power Cal dialog – Ref. Receiver"
"Power Cal dialog – Meas. Receiver"
"Power Cal dialog – Leveling Table"
A green checkmark indicates that the calibration data has been acquired successfully.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Apply
Is enabled when a new set of power calibration data has been acquired for either a
"Power" source or a "Meas. Receiver". The button applies all available source and/or
measurement receiver calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
Cal softtool