GUI reference
R&S
®
ZNA
550
User Manual 1178.6462.02 ─ 20
The value can be decreased down to 0.1
dB, but is only recommended for converters
with extremely steep power transfer characteristics. If
is checked, the
value is set to 1
dB and disabled.
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:STEP
ALC Step Size
If checked,
dB and the corresponding input field is dis-
abled. With this
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:POWer:ALC
Start Cal Sweep
Start the RF input power calibration sweeps for the selected port and settings and
close the dialog. See step
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling[:ACQuire]
Power Cal dialog – Source Flatness
In "Source Flatness" mode, the "Power Cal" dialog guides you through a source power
calibration of a single port.
The dialog shows the selected source port with its current cal power settings (see
Chapter 5.11.3.2, "Cal Power Config dialog"
A circuit diagram visualizes the calibration setup, which depend on the selected
●
"Ref. Receiver": connect a DUT or a Match to the related source port
●
"Power Meter": connect a power meter to the related source port
Access:
Background information
Refer to
Chapter 4.5.6.1, "Source power calibration"
"Combining SMARTerCal with scalar power calibration"
Cal softtool