Command reference
R&S
®
ZNA
1028
User Manual 1178.6462.02 ─ 20
Parameters:
<Boolean>
ON | OFF - Limit check on or off. A result is available even if the
limit check is disabled; see example for
*RST:
n/a (no ripple limit line defined after a reset)
Example:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
Manual operation:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit
<Limit>
Defines the ripple limit for ripple limit range no.
<Seg>
. A range must be created first to
"Rules for ripple test definition"
Tip:
To define several ripple limit ranges with a single command, use
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Number of the ripple limit range.
Parameters:
<Limit>
Ripple limit in the range. The unit is adjusted to the format of the
active trace (
Range:
Virtually no restriction for ripple limit ranges.
*RST:
n/a (no ripple limit line defined after a reset)
Default unit: See above.
Example:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
Manual operation:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
Returns the result of the ripple limit check in the previously defined limit range no.
<Seg>
. The response consists of two parameters:
●
<Boolean>
–
0 for "passed", 1 for "failed".
●
<Limit>
–
measured ripple in the limit range. A result is returned even if the limit
check in the range no. <Seg> is disabled; see example below.
A reset deletes all ripple limit ranges. Use
CALCulate<Ch>:RIPPle:FAIL?
to query
the result for global ripple limit check.