Command reference
R&S
®
ZNA
1127
User Manual 1178.6462.02 ─ 20
Note that these voltages are only applied while an RFFE command is executed using
CONTrol<Ch>:RFFE<Bus>:COMMand:SEND
or
.
Suffix:
<Ch>
.
Channel number
<Bus>
RFFE bus number
Parameters:
<Voltage>
Range:
0 to 2.5 V
Increment: 0.001 V
Default unit: V
Manual operation:
See
CONTrol<Ch>:RFFE<Bus>:TEST:DATA
<Test Data Voltage>
CONTrol<Ch>:RFFE<Bus>:TEST:CLOCk
<Test Clock Voltage>
CONTrol<Ch>:RFFE<Bus>:TEST:VIO
<OutputVoltage>
These commands define the output voltages for the voltage/current measurements on
the RFFE pins.
Note:
The voltages for DATA and CLOCk are always identical; their values cannot be
set independently.
The output voltages are applied using
, the volt-
age/current measurements are started using
.
Suffix:
<Ch>
.
Channel number
<Bus>
RFFE bus interface 1 or 2
Parameters:
<OutputVoltage>
Range:
0 V to 2.5 V
Increment: 1 mV
*RST:
0 V
Default unit: V
Example:
:CONTrol:RFFE1:TEST:DATA 1V
:CONTrol:RFFE1:TEST:CLOC?
DATA and CLOCK are set simultaneously, so this should return
1V
:CONTrol:RFFE1:TEST:VIO 1V
:CONTrol:RFFE2:TEST:CLOCk 2V
:CONTrol:RFFE2:TEST:DATA?
DATA and CLOCK are set simultaneously, so this should return
2V
:CONTrol:RFFE2:TEST:VIO 2V
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
SCPI command reference