SOURce<Ch>:POWer:CORRection:DATA
<Source>, <CorrData>...
SOURce<Ch>:POWer:CORRection:DATA:PORT<PhyPt>
<Source>, <CorrData>...
Reads or writes scalar source power correction data sets. A power correction data set
contains n real values where:
●
Each value corresponds to the ratio of the actual power at the reference plane
(value provided by the used source) to the uncalibrated power in dB.
●
The number n is equal to the number of sweep points.
Increasing (decreasing) the values in the correction data sets increases (decreases)
the power at the reference plane. Writing correction data (the setting command) fails if
the number of transferred values is not equal to the number of sweep points.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S
ZNA-K4).
Parameters:
<Source>
Identifier for the source of the calibrated wave:
'
A<n>
' or '
PORT<n>
' denote correction data for the analyzer port
no.
<n>
.
'
ESRC<n>
' or '
GEN<n>
' denote correction data for the external
generator no.
<n>
.
<CorrData>
Power correction values either in ASCII or block data format,
depending on the current
All numbers are interpreted as dB-values.
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source
port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-
separated real numbers.
SOUR:POW:CORR:DATA 'A1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.