GUI reference
R&S
®
ZNA
834
User Manual 1178.6462.02 ─ 20
Access:
:
●
System – [Setup] > "Frequency Converter" > "Leveling Dataset..."
●
Channel – [Cal] > "Use Cal" > "Active Power Cals..."
Background Information
See
Power Control
Select whether you want activate converter power control, if a leveling dataset is
.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:SLEVeling:STATe
Available
Shows the availability of a leveling data set.
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:STATe?
Load/Save
Use "Save" to persist the leveling dataset of the respective port to a
*.lev
file
(R&S
ZNA-proprietary, binary file format). Use "Load" to recall a leveling dataset from a
*.lev
file.
Remote command:
MMEMory:LOAD:CORRection:SLEVeling<PhyPt>
MMEMory:STORe:CORRection:SLEVeling<PhyPt>
Remove
Deletes the (global) leveling data set of the respective port.
Remote command:
SOURce:POWer<PhyPt>:CORRection:SLEVeling:DELete
Setup softtool