GUI reference
R&S
®
ZNA
439
User Manual 1178.6462.02 ─ 20
Response dialog
The "Response" dialog allows you to configure the limit line segment in detail.
Access:
> "Response" cells
For convenience, "Type", "Start Stimulus", "Stop Stimulus", and "Interpolation" are
repeated from the
of the "Define Limit Lines" dialog.
Linear/Formula
Allows you to decide how to define the line segment.
●
"Linear": Define the line segment as a straight line, connecting the endpoints
("Start Stimulus","Start Response") and ("Stop Stimulus","Stop Response").
In this case, you can also decide whether linear or logarithmic
●
"Formula": Define the line segment by a custom formula.
You can enter the formula in the text field to the right of the "Formula" button, or
Tap and hold the text field to open the
Remote command:
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula:STATe
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula
5.6.2
Ripple Test tab
Defines ripple limits for the measurement results, visualizes them in the diagrams and
activates/deactivates the ripple limit check.
A ripple test is a special type of limit test where the
difference
between response val-
ues in certain stimulus ranges must not exceed configurable limits (ripple limits).
Ripple limits can be defined for cartesian trace formats only and are limited to the trace
format they were configured for. If another format is selected, the ripple limit lines are
hidden and the limit check is temporarily disabled.
Background information
Refer to
Lines softtool