GUI reference
R&S
®
ZNA
695
User Manual 1178.6462.02 ─ 20
Note:
The loaded file is stored in the active recall set. Persisted recall sets contain the
full (de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:GLOop:DEEMbedding<group>
MMEMory:LOAD:VNETworks<Ch>:GLOop:EMBedding<group>
Set to Ideal Ground
This function is enabled as long as the 1-Port Data network is active. An imported S-
parameter set is replaced by the S-parameters of an ideal through connection, which
eliminates the transformation network.
Ground Loop per Port Group
If selected, each port group (= DUT) can have its own ground loop deembedding and
embedding networks. Otherwise the same de-/embedding network is used for all active
ports.
The checkbox is only visible if the instrument supports
, which is the case
for 4-port R&S
ZNA.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:GROup
5.14.9
Differential Match tab
Allows you to specify a 2-port embedding network for the differential mode of a bal-
anced port.
Such a network is either defined
●
via its S-parameter stored in a two-port Touchstone file (
*.s2p
) or
●
by specifying the parameters of a "Shunt L, Shunt C" lumped element model
Use the complementary dock widget to access the balanced port configuration and to
activate or deactivate "Differential Match" embedding for selected balanced ports (see
Chapter 5.14.1.8, "Differential Match panel"
Background information
Refer to
Chapter 4.6.2.8, "Differential match embedding"
Offset Embed softtool