Command reference
R&S
®
ZNA
1094
User Manual 1178.6462.02 ─ 20
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:DELete
Deletes all port sets (including port pairs) previously defined for embedding.
Suffix:
<Ch>
.
Channel number
<ListId>
This suffix is ignored
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:PPAir:
Usage:
Event
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:PARameters:
C<1|2|3>
<CircuitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:PARameters:
C<1|2|3>?
<CircuitModel>
Specifies the capacitance value C<i> in the different lumped circuit models for port pair
embedding.
In the query form, the <Capacitance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId>
Index of the affected port pair (see
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine
<1|2|3>
Index i of the capacitance C<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<Capacitance>
Capacitance Ci
Range:
-1 mF to 1 mF
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel>
STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS |
STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Circuit model whose capacitance C<i> shall be set, see
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:PPAir:
Manual operation:
See
SCPI command reference