Command reference
R&S
®
ZNA
1363
User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Channel number
Parameters:
<arg0>
Physical port number
Instead of port numbers 1, 2, 3 and 4 you can also use the con-
stants ARECeiver, BRECeiver, CRECeiver and DRECeiver,
respectively.
Default unit: n/a
<arg1>
Attenuation factor for the received wave.
Range:
0 dB, 10 dB, 20 dB, 30 dB. The analyzer rounds
any entered value below the maximum attenuation
to the closest step.
*RST:
0 dB
Default unit: dB
Example:
POW:ATT AREC, 10
Set an attenuation factor of 10 dB for the waves received at test
port 1 and channel no. 1. The other test ports and channels are
not affected.
Manual operation:
See
[SENSe<Ch>:]POWer:GAINcontrol
<ReceiverName>, <Mode>
[SENSe<Ch>:]POWer:GAINcontrol?
<ReceiverName>
Defines port-specific gain settings for the related channel.
Note:
●
This applies to the wideband signal path only.
●
These settings will only take effect if
[SENSe<Ch>:]POWer:GAINcontrol:
MANual
for channel <Ch>.
●
If you also want to distinguish between different sweep segments, use
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
Suffix:
<Ch>
.
Channel number
Parameters:
<Mode>
LDIStortion | LNOise
LDIStortion
0 dB gain, for high input levels
LNOise
10 dB gain, for low input levels
*RST:
LDIStortion
Parameters for setting and query:
<ReceiverName>
String parameter defining the wave(s) followed by the drive port;
see examples below and
SCPI command reference