Command reference
R&S
®
ZNA
1248
User Manual 1178.6462.02 ─ 20
<KitLabel>
String parameter containing the label of the calibration kit, usu-
ally the serial number.
Setting parameters:
<Mode>
DELay | ELENgth
DELay – Keysight modelling
ELENgth – ZVR compatible modelling
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:LABel
<CalKitName>[, <KitLabel>]
Assigns a label to an imported or user-defined calibration kit.
Parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
<KitLabel>
String parameter containing the calibration kit label.
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Create a new cal kit "New Kit 1" and assign an open (f) standard
for the N 50 Ω connector type with specific properties.
CORR:CKIT:LAB 'New Kit 1', 'Test kit created
today'
Label the previously created kit.
CORR:CKIT:LAB? 'New Kit 1'
Check the label.
CORR:CKIT:DEL 'New Kit 1'
Delete the kit.
[SENSe:]CORRection:CKIT:SELect
<ConnectorType>[, <CalKitName>]
Selects the calibration kit to be used, specifying its connector type and name
(optional).
Tip:
The command is suitable for connector types with arbitrary, user-defined names.
For standard connector types you can use the command
.
Parameters:
<ConnectorType>
Connector type, e.g. a user-defined connector type (string vari-
able).
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer.
SCPI command reference