Command reference
R&S
®
ZNA
1427
User Manual 1178.6462.02 ─ 20
Parameters:
<arg0>
The multiplication factor to calculate the minimum frequency off-
set.
To get the resulting frequency offset, the minimum frequency off-
set is rounded to a multiple of the current frequency step size.
Range:
1 to 100
*RST:
10
Example:
See
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:
Manual operation:
See
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition
This command is used to query the operation mode of Simultaneous Measurement
with Frequency Offset. The list below contains all possible return values of this query.
Please note that this query will return an error if the channel specified by <Ch> does
not contain any trace. This is due to the fact that without traces no measurements are
performed so that no information about Simultaneous Measurement with Frequency
Offset is available.
This query waits until all previous commands are fully executed because these com-
mands may affect the operation mode of Simultaneous Measurement with Frequency
Offset.
Suffix:
<Ch>
.
Channel number.
If unspecified, this is set to 1.
<Grp>
Return values:
<errorCode>
0, "switched off"
Simultaneous Measurement with Frequency Offset is switched
OFF, meaning
SOURce<Ch>:GROup<Grp>:SIMultaneous:
1, "active"
Simultaneous Measurement with Frequency Offset is active,
meaning it has been activated via
SOURce<Ch>:GROup:SIMultaneous:FOFFset:STATe ON
and all preconditions for this mode to be operational are met.
-1, "too few port groups"
There must be at least two port groups defined, otherwise no
parallel measurement is possible.
-2, "invalid port groups"
The port groups have to be set up exactly as in the VNA GUI,
when "Enabled" is selected in the "Define Parallel Measure-
ment" dialog.
SCPI command reference