GUI reference
R&S
®
ZNA
504
User Manual 1178.6462.02 ─ 20
When "Time" is checked, two new columns appear in the table. The first column is
titled "Segment Time" or "Meas Delay", depending on the selected radio button
below the "Time" checkbox. The second column is titled "Auto" and is used to acti-
vate automatic sweep time setting.
"Segment Time" is the total measurement time for the sweep segment. The mini-
mum segment sweep time to be set is equal to the estimated value in "Auto" mode.
"Meas Delay" sets a delay time allowing the DUT to settle before the hardware set-
tings of the analyzer are changed and a new partial measurement is started.
Changing the "Meas Delay" modifies the "Segment Time" and vice versa.
"Auto" minimizes the sweep time. If "Auto" is selected for a segment, the columns
"Segment Time" or "Meas Delay" (in the "Define Segments" dialog cannot be edi-
ted). "Segment Time" indicates the estimated sweep time, depending on the cur-
rent measurement settings, the "Meas Delay" is 0 s. The segment sweep time and
point delay values are maintained until changed explicitly if "Auto" is switched off.
Remote command:
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:DEFine
[SENSe<Ch>:]SEGMent<Seg>:DEFine:SELect
[SENSe<Ch>:]SEGMent<Seg>:INSert
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>[:STATe]
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>:POWer[:LEVel]
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:SUM?
OUTPut<Ch>:UPORt:SEGMent<Seg>:STATe
OUTPut<Ch>:UPORt:SEGMent<Seg>[:VALue]
CONTrol:AUXiliary:C[:DATA]
5.10.3
Trigger tab
Selects the source of the trigger signal and provides additional trigger settings.
Trigger system of the analyzer
The trigger system is used to synchronize the analyzer's actions with events that can
be provided by an internal or external signal or user-generated ("Manual Trigger"). Trig-
gered measurements are an alternative to the default mode ("FreeRun", "Continuous"
sweep), where the measurement is continuously repeated without fixed time reference.
Sweep Softtool