In the converter configuration above, P1 and P2 drive the RF IN ports of converter 1
and 2, respectively, and the Converter LO port drives the LO IN ports of both convert-
ers.
If the source flatness calibration at the converter port is enabled, then the flatness cal
at the RF IN port is disabled (and grayed out). However, the source flatness cals at the
converter RF in or out ports are only used, if the converter port is
not
configured to use
.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe]
[SENSe<Ch>:]CORRection:PCAL
5.11.4.3
Calibration Manager dialog
The "Calibration Manager" dialog stores user correction data to the cal pool and
assigns stored correction data to channels.
See
Chapter 4.5.3, "Calibration pool"
on page 178 for background information.
Access:
Channel – [Cal] > "Use Cal" > "Cal Manager..."
Cal softtool