GUI reference
R&S
®
ZNA
668
User Manual 1178.6462.02 ─ 20
Overview
See
Port
Physical test port of the analyzer. You can define independent loss parameters for all
ports.
Remote command:
The
<PhyPt>
numeric suffix in the
[SENSe<Ch>:]CORRection:...
commands
identifies the physical port.
Loss at DC / Loss at 1st Freq / 1st Freq for Loss
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd Freq / Loss at 2nd
Auto Length and Loss
Determines the length offset and one-way loss parameters for the receive port of the
active trace. See
Chapter 4.6.1.4, "Auto Length and Loss"
Note:
●
If "Auto Length and Loss" is used with a line connected to a test port, the end of the
line should be left open.
●
To calculate the electrical length offset and loss of port
i
, you have to select a trace
that uses port
i
as one of its receive ports. It is
not
relevant for the "Auto Length
If the measured quantity is a ratio, or if it is derived from a ratio, its receiving port is
given as the index of the wave quantity in the numerator. If the active trace shows an
S-parameter S
ij
, then "Auto Length" adds a length offset at port i. If the receive port is
balanced, then the same offset and length is applied to both its physical ports.
Remote command:
[SENSe<Ch>:]CORRection:LOSS:AUTO
Offset Embed softtool