Command reference
R&S
®
ZNA
1312
User Manual 1178.6462.02 ─ 20
[SENSe:]CORRection:EDELay:VNETwork
<Boolean>
Changes the position of the offset calculation in the "Offset Embed" calculation chain.
Parameters:
<Boolean>
OFF (0)
The offset is calculated
before
de-/embedding (default).
ON (1)
The offset is calculated
after
de-/embedding.
*RST:
0
Manual operation:
See
"Offset > Calculate after De-/Embed."
[SENSe<Ch>:]CORRection:EWAVe[:STATe]
<Boolean>
Activates or deactivates the wave correction logic for wave quantities and ratios.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
*RST:
ON (1)
Manual operation:
See
[SENSe<Ch>:]CORRection:LOSS<PhyPt>
<LossAtFrequency>
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:SECond
<LossAtFrequency>
These commands define the
f
1
and
f
2
.
The reference frequencies can be set using
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<LossAtFrequency> Loss at reference frequency
Range:
-200 dB to +200 dB
Increment: 0.001 dB
*RST:
0 dB
Default unit: dB
Manual operation:
See
"Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS:AUTO
<Activate>
) function to the active trace of channel <Ch>.
SCPI command reference