Concepts and features
R&S
®
ZNA
133
User Manual 1178.6462.02 ─ 20
Examples:
●
"b
2
/a
1
Src Port 1" is the ratio of the outgoing wave at DUT port 2 and the incident
wave at DUT port 1 (i.e. DUT port 1 ist stimulated). This corresponds to the forward
transmission coefficient S
21
.
●
"b
1
/a
1
Src Port 1" is the ratio of the outgoing wave at DUT port 1 and the incident
wave at DUT port 1 (i.e. DUT port 1 ist stimulated). This corresponds to the forward
reflection coefficient S
11
.
4.3.5.3
Detector settings
The "Detector" settings select the algorithm that is used to calculate the displayed
measurement points from the raw data. The "Detector" can be selected in the More
Params, "More Wave Quantities", and More Ratios dialogs.
Figure 4-6: More Wave Quantities dialog
The following detectors are available:
●
"Normal"
selects the default detector mode where each valid measurement point
is displayed without modification. The analyzer then proceeds to the next sweep
Measurement results