Command reference
R&S
®
ZNA
939
User Manual 1178.6462.02 ─ 20
OFF (0)
Display all traces in a single diagram.
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:DELT:TDR
<Boolean>
If the
option R&S
ZNA-K2 is installed, this command allows you
to add the TDR-based converted impedance traces to the recommended trace configu-
ration.
Parameters:
<Boolean>
OFF (0)
TDR-based impedance traces added
ON (1)
TDR-based impedance traces not added
*RST:
OFF
Options:
R&S
ZNA-K231
Manual operation:
See
CALCulate:FMODel:EZD...
Commands for
Eazy Deembedding
Chapter 4.7.14, "Eazy de-embedding based
CALCulate:FMODel:EZD<Ph_pt>:COUPon:MEASure
For a fixture modeling with the Easy De-embedding (EZD) tool, this command starts
the measurement of the symmetrical 2xThru coupon at the active ports (see
CALCulate:FMODel:EZD<Ph_pt>:COUPon[:STATe]
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage:
Event
Options:
R&S
ZNA-K210
Manual operation:
See
"Measure / Measure Open / Measure Short"
CALCulate:FMODel:EZD<Ph_pt>:COUPon:MEASure:FILename
<String>
For a fixture modeling with the EZD tool, this command loads the coupon properties
from a Touchstone file.
Suffix:
<Ph_pt>
.
This suffix is ignored.
SCPI command reference