GUI reference
R&S
®
ZNA
572
User Manual 1178.6462.02 ─ 20
5.11.3
Pwr Cal Settings tab
Gives access to all functions for power meter and power calibration data handling
(transmission coefficients). Power calibration of the R&S
ZNA is a fully menu-guided
process.
Efficient power calibration procedure
●
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal" –
"Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Pwr Cal Settings" tab.
●
Select "Cal Power..." if you use an amplifier between the source port and the DUT.
●
Select "Transm. Coefficients..." if you want to modify the power calibration proce-
dure.
Background information
Refer to
Chapter 4.5.6, "Scalar power calibration"
5.11.3.1
Controls on the Pwr Cal Settings tab
Cal softtool