"Power Meter
only"
Use a power meter for the calibration of the reference receiver
and
the flatness calibration.
"Ref. Receiver
only"
Perform a flatness calibration using the reference receiver. Do not
recalibrate the reference receiver; use the best available reference
receiver calibration instead.
For external generators and the optional
on page 275 port, a
power meter must be used.
"Ref. Receiver after" is the default behavior.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:METHod
Embedding
Defines whether the source flatness cal uses the configured offset de-/embedding to
get the power values at the DUT (instead of the calibratied reference plane).
This setting applies to all ports and channels. Default is "Off" (no offset or de-/embed-
ding used).
Remote command:
SOURce:POWer:CORRection:COLLect:VNETworks
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the power calibration of a source port. The last configured power meter is selected
by default. See
Remote command:
SOURce:POWer:CORRection:PMETer:ID
Auto Zero
Initiates an automatic zeroing procedure of the selected power meter.
The power meter must be disconnected from the RF power; see
on page 287. A message indicates that zeroing is finished.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
5.11.3.2
Cal Power Config dialog
The "Cal Power Config" dialog adjusts the target power for the power calibration (cal
power) and defines the target power for the reference receiver calibration. These set-
tings are particularly important for test setups involving external attenuators or amplifi-
ers.
Access:
Channel – [Cal] > "Pwr Cal Settings" > "Cal Power..."
Related information
The settings in the "Cal Power Config" dialog are also used to define port-specific
source powers in arbitrary mode; refer to
Cal softtool