Command reference
R&S
®
ZNA
985
User Manual 1178.6462.02 ─ 20
Parameters:
<OutFormat>
MLINear | MLOGarithmic | PHASe | POLar | GDELay | REAL |
IMAGinary | SWR | LINPhase | LOGPhase | IMPedance |
ADMittance | DEFault | COMPlex | MDB | MLPHase |
MDPHase | MIMPedance | PIMPedance | PADMittance |
MADMittance | MPIMpedance | MPADmittance | INDX | NOISe
DEFault means the default marker format is dynamically adjus-
ted to the selected trace format (
For the other marker formats see the table below and the
description in
*RST:
DEFault
Manual operation:
See
SCPI
GUI
MLINear
"Lin Mag"
MLOGarithmic
MDB (for R&S
ZVR compatibility)
"dB Mag"
PHASe
"Phase"
POLar
COMPlex (for R&S
ZVR compatibility)
"Real Imag"
GDELay
"Delay"
REAL
"Real"
IMAGinary
"Imag"
SWR
"SWR"
LINPhase
MLPhase (for R&S
ZVR compatibility)
"Lin Mag Phas"e
LOGPhase
MDPhase (for R&S
ZVR compatibility)
"dB Mag Phase"
IMPedance
"R + j X" for reflection measurement traces
"R + j X series" for transmission measurement
traces
PIMPedance
"R + j X", for reflection measurement traces
"R + j X parallel" for transmission measurement
traces
ADMittance
"G + j B" for reflection measurement traces
"G + j B series" for transmission measurement
traces
PADMittance
"G + j B" for reflection measurement traces
"G + j B parallel" for transmission measurement
traces
SCPI command reference