Command reference
R&S
®
ZNA
1129
User Manual 1178.6462.02 ─ 20
Manual operation:
See
2 μA
20 μA
200 μA
2 mA
20 mA
Shunt
10 Ω
100 Ω
1 kΩ
10 kΩ
100 kΩ
CONTrol:RFFE<Bus>:TEST:DATA:VOLTage?
CONTrol:RFFE<Bus>:TEST:CLOCk:VOLTage?
CONTrol:RFFE<Bus>:TEST:VIO:VOLTage?
Returns the results of the voltage measurement on the related RFFE pin.
Suffix:
<Bus>
.
RFFE bus
Return values:
<Measured>
Measured voltage
Usage:
Query only
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol:RFFE<Bus>:TEST:DATA:CURRent?
CONTrol:RFFE<Bus>:TEST:CLOCk:CURRent?
CONTrol:RFFE<Bus>:TEST:VIO:CURRent?
Returns the results of the current measurement on the related RFFE pin.
Suffix:
<Bus>
.
RFFE bus
Return values:
<Measured>
Measured current
Usage:
Query only
Options:
R&S ZN-B15/Z15 Var. 03
Manual operation:
See
CONTrol<Ch>:SEGMent<Nr>:SEQuence:COUNt?
For segmented sweeps this command queries the length of the command/switch
sequence defined for a particular channel and sweep segment.
The command/switch sequences are defined using
CONTrol<Ch>:SEGMent<Nr>:GPIO<Port>:
.
Suffix:
<Ch>
.
Channel number
<Nr>
Segment number
SCPI command reference