Command reference
R&S
®
ZNA
1518
User Manual 1178.6462.02 ─ 20
ON (1)
New traces are created with
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe]
set to
ON
(1)
.
*RST:
Not affected by *RST; factory default is OFF
Manual operation:
See
SYSTem:USER:DISPlay:TITLe
<String>
Defines a title for the remote display, i.e. for the screen that is shown at the instrument,
if a remote session is established and the standard display is switched off.
Parameters:
<String>
Title string.
A \n in the string starts a new line; see example.
*RST:
empty string
Example:
SYST:USER:DISP:TITL "DON'T TOUCH\nRemote test
running..."
Define a title for the remote display.
SYSTem:USER:KEY
<Key>[, <Label>]
Labels a user-defined key in the remote display. In the query form the command
returns whether or not a user-defined key was tapped or clicked.
Parameters:
<Key>
Number of the user key
0
–
Delete all user keys and restore the default keys ("Go to
Local", "Display Off").
1 to 8
–
User key numbers
Range:
0 to 8
*RST:
0
<Label>
Label for user key no. 1 to 8 (string variable)
SCPI command reference