Switching on the display is ideal for program test purposes but tends to slow down the
measurement. Therefore, it is recommended to switch off the display in real measure-
ment applications where a tested program script is to be executed repeatedly.
The analyzer provides a third display option where the measurement screen is only
updated when triggered by the remote control command
SYSTem:DISPlay:UPDate
ONCE
.
The instrument remains in the remote state until it is reset to the manual state via the
GUI or via remote control (see
Chapter 6.1.4.2, "Returning to manual operation"
on page 851). You can also lock the remote (touch) screen using
SYSTem:TSLock
SCReen
.
A tooltip across the bottom of the remote screen indicates a remote command error.
You can switch off this tooltip using
SYSTem:ERRor:DISPlay:REMote OFF
.
SCPI commands: