GUI reference
R&S
®
ZNA
511
User Manual 1178.6462.02 ─ 20
5.10.4.1
Controls on the Sweep Control tab
Continuous / Single
Activate either continuous (default) or single sweep mode.
●
In "Continuous" sweep mode, the analyzer measures the related channel (or chan-
nels) continuously, repeating the current sweep over and over.
●
In "Single" sweep mode, the measurement is stopped after the configured number
of
.
initiates a new measurement cycle.
For the DEFAULT
, this setting applies to the active channel only.
Use
All Channels Continuous/All Channels on Hold
to select a common sweep mode
for all channels.
For remote languages other than DEFAULT, this setting applies to all channels.
Remote command:
See also:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
CONFigure:CHANnel:MEASure:ALL[:STATe]
Controlled Timing
Activates/deactivates the controlled timing mode, which is provided by
ZNA-K28.
Note:
This checkbox is only available if option R&S
ZNA-K28 is installed. It is only
enabled if the current setup contains a single channel.
Remote command:
Sweep Softtool