Command reference
R&S
®
ZNA
1308
User Manual 1178.6462.02 ─ 20
TYPE
Calibration type (see <Type> parameter in
]CORRection:COLLect[:ACQuire]:SELected
PORTs
The related test port numbers (comma-separated list of integers)
SPORt
For SMARTer calibrations this returns the port to which a power-
meter was connected (0 otherwise)
THRoughs
Measued Throughs: comma-separated list of ordered port num-
ber pairs <pno1-pno2>
TSTamp
Timestamp (in UTC)
LTSTamp
Timestamp in local time
CKIT
For each port, the name of the used calibration kit is stored with
the calibration.
The query can return one of the following:
–
the cal kit name
(string)
if available and unique
–
'Multiple'
if cal kit names are available but not unique, i.e. if <PhyPt> was
not specified and multiple cal kits were used (multi-port cals with
different connector types, merged cals)
<Index>
Currently not used
Example:
SENSe:CORRection:DATA:PARameter?
Unrestricted query. Result looks like this:
100000,8500000000,201,-10,LIN
SENSe:CORRection:DATA:PARameter? STARt
...
SENSe:CORRection:DATA:PARameter? STYPe
Query settings one by one.
SENSe:CORRection:DATA:PARameter? RATTenuation
Query receiver attenuations. Result looks like this:
1,0.000000,2,0.000000
Usage:
Query only
Manual operation:
See
"Ch<n> Calibration Properties/Cal Group <n> Properties"
[SENSe<Ch>:]CORRection:DATA:PARameter:COUNt?
Gets the number of active system error corrections for channel <Ch>.
SCPI command reference