Command reference
R&S
®
ZNA
1338
User Manual 1178.6462.02 ─ 20
Manual operation:
See
[SENSe<Ch>:]FREQuency:IMODulation:ORDer<Im>[:STATe]
<Boolean>
Enables or disables the measurement of the intermodulation products of order <IM
order>.
Suffix:
<Ch>
.
Channel number
<Im>
Order of IM products
Parameters:
<Boolean>
ON | OFF – enables or disables measurement
*RST:
ON (for <Im> = 3), OFF (for <Im> > 3)
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver
<PortNumber>
Selects the receiving port for the intermodulation measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<PortNumber>
Analyzer port number
Range:
1 to the number of analyzer ports
*RST:
2
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
Manual operation:
See
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum:MORDer
<IMOrder>
Defines the maximum order of intermodulation products for the intermodulation spec-
trum measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<IMOrder>
Maximum order of IM products
Range:
3 | 5 | 7 | 9
*RST:
3
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:
Manual operation:
See
SCPI command reference