Command reference
R&S
®
ZNA
1300
User Manual 1178.6462.02 ─ 20
Parameter
Type ("Calibration Presetting" dialog)
CPUOsm
PUOSM Scalar Vector Mixer
MTOSm | MUOSm | MOSM | MTRL
of type OSM, UOSM, OSM, or TRL
[SENSe<Ch>:]CORRection:COLLect:NFIGure:END
Terminates a noise figure calibration, discarding the acquired calibration data.
A noise figure calibration is initiated using
[SENSe<Ch>:]CORRection:COLLect:
and data are aquired using
. It must be terminated by one of the commands
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE
or
[SENSe<Ch>:]CORRection:COLLect:NFIGure:END
.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
All commands related to a noise figure calibration must use the
same channel number.
Example:
See
Chapter 8.2, "Condensed programming examples"
Usage:
Event
Options:
R&S
ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE
Completes a noise figure calibration, storing and applying the acquired calibration data.
A noise figure calibration is initiated using
[SENSe<Ch>:]CORRection:COLLect:
and data are aquired using
. It must be terminated by one of the commands
[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE
or
]CORRection:COLLect:NFIGure:END
.
In case a noise measurement is missing, an error is reported. In this case, simply per-
form the missing ACQ step, then repeat SAVE.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
All commands related to a noise figure calibration must use the
same channel number.
Example:
See
Chapter 8.2, "Condensed programming examples"
Usage:
Event
Options:
R&S