Command reference
R&S
®
ZNA
1301
User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt
<arg0>, <arg1>, <arg2>,
<arg3>, <arg4>
Initiates a noise figure calibration for the related channel.
The calibration sweeps are started using
[SENSe<Ch>:]CORRection:COLLect:
, the noise figure calibration is terminated using
]CORRection:COLLect:NFIGure:SAVE
or
.
Suffix:
<Ch>
.
Number of the channel to be noise figure calibrated.
The subsequent noise figure calibration commands must use the
same channel number.
Setting parameters:
<arg0>
Number of the physical VNA port connected to the DUT output.
<arg1>
Number of the physical VNA port connected to the DUT input.
<arg2>
ON | OFF - Boolean parameter, indicating whether an external
attenuator is used.
<arg3>
Attenuation of the source level during the source noise calibra-
tion.
<arg4>
Attenuation of the source level during the measurement (i.e.
after calibration).
Example:
See
Chapter 8.2, "Condensed programming examples"
Usage:
Setting only
Options:
R&S
ZNA-K30
[SENSe<Ch>:]CORRection:COLLect:NFIGure[:ACQuire]
<arg0>
Starts one of the three noise calibration sweeps that are required for a noise figure cali-
bration of the related channel.
The noise figure calibration is initiated using
[SENSe<Ch>:]CORRection:COLLect:
and must be terminated by one of the commands
]CORRection:COLLect:NFIGure:END
or
.
Note:
A noise figure calibration must be terminated with one of the following com-
mands, to be sent immediately after the sequence of
...[:ACQuire]
commands:
●
Abort the noise figure calibration, discarding the acquired calibration data.
●
Terminate the noise figure calibration, applying the acquired calibration data (the
active channel is calibrated) These commands ensure that the analyzer is able to
start a new measurement sweep.
SCPI command reference