Command reference
R&S
®
ZNA
895
User Manual 1178.6462.02 ─ 20
<FwCount>
Number of first sweep to be read. 1 denotes the first sweep
acquired, 2 denotes the second and so forth. The sweep count
in single sweep mode is defined via
Range:
1 to sweep count
<FwCountEnd>
Number of last sweep to be read. If this parameter is omitted, it
is implicitly set to
<FwCount>
(a single sweep is read).
Range:
<FwCount> to sweep count
Return values:
<Data>
Response values either in ASCII or block data format, depend-
ing on the current
Example:
SWE:COUN 10
Define the number of sweeps (10) to be measured in single
sweep mode.
INIT:CONT OFF; :INIT;
Activate single sweep mode and start a single sweep sequence
in channel no. 1. No synchronization is necessary.
if (CALC:DATA:NSW:COUN? > 4)
CALC:DATA:NSW:FIRS? SDAT, 5
Wait until 5 sweeps have been measured, then query the results
of the 5
th
sweep.
Chapter 8.2.4.3, "Retrieving the results of previous
Usage:
Query only
CALCulate<Chn>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
Usage:
Query only
7.3.1.4
CALCulate:DLINe...
The
CALCulate:DLINe...
commands control the horizontal line used to mark and
retrieve response values (display line).
.................................................................................................895
.......................................................................................896
CALCulate<Chn>:DLINe
<Position>
Defines the position (response value) of the horizontal line.
SCPI command reference