background image

CO

2

 Analyzer

2810/2815/2820

Instruction 19-9222

Operation & Maintenance 

Rev. 9 – November 2005

Product Leadership • Training • Service • Reliability

2

3

Содержание 2810

Страница 1: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 Operation Maintenance Rev 9 November 2005 Product Leadership Training Service Reliability 2 3...

Страница 2: ...ING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGA TION LIABILITY RIGHT CLAIM OR REMEDY IN CONTRACT OR TORT WHETHER OR NOT ARISING...

Страница 3: ...s Level Tone 2815 2820 5 2 11 Fault Condition Warning 5 2 12 Powering Analyzer from Charger 6 2 13 Troubleshooting for Slow or Low Readings 2820 6 2 14 Testing Incubators 2820 7 3 0 MAINTENANCE 8 3 1...

Страница 4: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 ii Notes...

Страница 5: ...20 also includes a desiccant filter to be used when sampling areas with high humidity All analyzers can store gas readings in memory which can later be downloaded to a personal computer via its integr...

Страница 6: ...t least 3 seconds or until the display goes blank When first switched on there is a warm up period be fore the CO2 level is displayed approximately 60 seconds for the 2810 2815 analyzer and 20 seconds...

Страница 7: ...ions where longer sampling lengths are required or where it is neces sary to draw against a vacuum 2 4 Pump Contamination 2815 2820 Over time the pump can become contaminated leading to a slower respo...

Страница 8: ...g of the readings at a preset interval factory set at 30 seconds The analyzer can store up to 100 snapshot readings and 800 sets of continuous data logged readings Once the storage capacity has been e...

Страница 9: ...for 5 seconds and then automatically turns off 2 10 Audible Gas Level Tone 2815 2820 An audible tone which will increase in pitch with increasing CO2 levels is activated by pressing and holding down...

Страница 10: ...ooting for Slow or Low Readings 2820 It normally takes between 1 and 2 minutes to obtain a peak reading when testing an incubator A slow or low reading can be caused by either a clogged desiccant filt...

Страница 11: ...ovided with the instrument as an accessory to prevent water from be ing drawn into the instrument The water trap is installed between the incubator and the Nafion tubing as shown in Figure 2 The need...

Страница 12: ...the unit will invalidate the warranty 3 4 Software Version Serial Number With the analyzer switched OFF and while holding down the f1 button switch ON the analyzer to display its software version and...

Страница 13: ...tery should normally be charged at room temperature Charging at temperatures below 54 F 12 C should be avoided since this may cause a false indication of when the battery is charged and could also dam...

Страница 14: ...Calibration be per formed at the beginning of each work day or each time the instrument is switched OFF ON The analyzer should also be checked on a periodic basis with a higher concentration of CO2 to...

Страница 15: ...ing fresh ambient air If the air calibration procedure was successful the display will show AirCal OK If the procedure was unsuccessful the message AirCal Failed will be displayed If this happens retr...

Страница 16: ...e analyzer automatically switches to its Air Gas screen before proceeding f1 Air f3 Gas 2 Press the f3 button to display the Gas Calibration screen f1 2500 340 The top line shows the preset calibratio...

Страница 17: ...level displayed on the second line of the Gas Calibration screen should slowly rise to the value displayed in the top line 5 Allow the calibration gas to flow until the second line of the Gas Cali bra...

Страница 18: ...d hold down the f2 button until the User Cal screen is dis played Note that battery status is first displayed for approx imately 2 seconds User Cal Wait until the analyzer automatically switches to it...

Страница 19: ...5 This completes the N2 Calibration procedure If desired the analyzer can now be spanned to a known concentration of CO2 gas per Section 3 7 2 2 3 7 2 2 Gas Calibration WARNING This procedure assumes...

Страница 20: ...hat is close to the gas level being monitored This operation requires that the preset calibration gas value be reset higher or lower using the optional BACH COM software version 2 22 or higher 4 Remov...

Страница 21: ...ect concentration of calibration gas If the procedure is still unsuccessful then the analyzer must be returned to a Bacharach Service Center for evaluation 7 This completes the Gas Calibration procedu...

Страница 22: ...2 Luer Fitting Male 2815 2820 103 2563 Gas Sample Tube 6 inches 2815 11 0118 Particulate Filter 2815 2820 54 0548 Nafion Portex Tubing Kit 2820 19 3318 Water Trap 2820 19 3265 Accessories 12 VDC Charg...

Страница 23: ...a Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xico Playa Regatas No 473 Tercer Piso Col Militar Marte Delegaci n Iztacalco 08830 M xico D F M xico Phones 52 555...

Страница 24: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 20 Notes...

Страница 25: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 21 Notes...

Страница 26: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 22 Notes...

Страница 27: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 23 Notes...

Страница 28: ...Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 800 736 4666 Website www bacharach inc com E mail help bacharach inc com Printed in U S A...