Analyzer 2810/2815/2820
Instruction 19-9222
16
Wat untl the analyzer automatcally swtches to ts ‘N2/Gas’ screen
before proceedng.
f1-> N2
f3-> Gas
3. Press the
f
3
button to dsplay the Gas Calbraton screen.
f1->05.0
00.0
The top lne shows the preset calbraton gas value that must be ap-
pled to the analyzer’s nlet. The bottom lne shows the current CO
2
readng.
Note:
The analyzer is most accurate around the calibration
gas value. If the analyzer, however, is routinely used to moni-
tor CO
2
levels that are much higher or lower than the preset
value, then it is recommended that the analyzer be calibrated
on a concentration of gas that is close to the gas level being
monitored. This operation requires that the preset calibration
gas value be reset higher or lower using the optional BACH-
COM software, version 2.22 or higher.
4. Remove the 100% N
2
cylnder (f used); then attach a CO
2
calbraton
gas cylnder to the nlet of the analyzer as shown n Fgure 3. The ap-
pled CO
2
concentraton must equal the gas value dsplayed n the
top
lne of the Gas Calbraton screen.
Important:
Do not attempt to calibrate the analyzer if the
pressure gauge closest to the tank reads below 5 psi.
5. Adjust the regulator attached to the gas cylnder for an outlet pressure
of 2 psi (equivalent to a flow rate of approximately 120 cc/minute). Ob
-
serve that the gas level dsplayed on the second lne of the Gas Calbra-
ton screen should slowly rse to the value dsplayed n the top lne.
Содержание 2810
Страница 4: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 ii Notes...
Страница 24: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 20 Notes...
Страница 25: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 21 Notes...
Страница 26: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 22 Notes...
Страница 27: ...CO2 Analyzer 2810 2815 2820 Instruction 19 9222 23 Notes...