Command reference
R&S
®
ZNA
1191
User Manual 1178.6462.02 ─ 20
MMEMory:LOAD:CORRection:TCOefficient<Ch>
<TraceFile>[, <Trace>]
Loads power meter (two-port) transmission coefficients from the specified power meter
correction file or trace file to channel <Ch>.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceFile>
String parameter specifying the name and directory of the loa-
ded trace file. The R&S
ZNA supports power meter correction
list files (
*.pmcl
*csv
, and Touchstone
(
*.s1p, *.s2p, ...
) files.
If no path is specified the analyzer searches the current direc-
tory, to be queried with
The file extensions
*.s<n>p
,
*csv
, and
*.pmcl
for Touch-
stone, ASCII, and power meter correction list files are manda-
tory.
*RST:
n/a
<Trace>
Optional string parameter: For multiport Touchstone files
(
*.snp
,
n > 1
), the parameter refers to a particular S-parame-
ter trace (
'S11', 'S12', ...
). For ASCII (
*.csv
) files, the
parameter references a trace name in the file (case sensitive). If
the parameter is omitted, the first trace in the specified file is
imported.
This parameter is not used for power meter correction list files
(
*.pmcl
).
*RST:
n/a
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
Manual operation:
See
MMEMory:LOAD:EYE:BPATtern
<TraceName>, <TraceFile>
Loads a user-defined bit pattern for the related eye diagram from a 7-bit ASCII file.
The pattern is repeated until the configured length is reached (see
Loading a pattern from file implicitly sets the type of bit stream to USER (see
CALCulate<Chn>:EYE:INPut:BPATtern:TYPE
).
Setting parameters:
<TraceName>
Name of the related eye diagram
SCPI command reference