GUI reference
R&S
®
ZNA
320
User Manual 1178.6462.02 ─ 20
●
"b1/a2 Source Port 2" is the ratio of the wave quantities b
1
and a
2
and corresponds
to the S-parameter S
12
(reverse transmission coefficient).
The analyzer can also measure arbitrary ratios for other source ports; see
ter 5.2.4.2, "More Ratios dialog"
Tip:
In the trace list, the source port is indicated in brackets. "b2/a1(P1)" denotes the
ratio b
2
/a
1
with source port 1.
Remote command:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "B2/A1" | ...
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "B2/A1" | ...
Wave Correction
Determines whether wave quantities and ratios are system error corrected in the active
channel.
If checked (default), the "best available" calibration is applied, with the following prece-
dence:
●
Scalar power calibration
●
User system error correction
●
Factory system error correction
Note:
●
Only a single calibration is applied.
●
If "Wave Correction" is turned off, no calibration is applied for these measurement
quantities at all.
Remote command:
[SENSe<Ch>:]CORRection:EWAVe[:STATe]
5.2.4.2
More Ratios dialog
The More Ratios dialog allows you to select arbitrary ratios between wave quantities b
i
,
a
j
as measured quantity. The ratios can be calculated with arbitrary source port and dif-
ferent detector settings.
Access:
Trace – [Meas] > "Ratio" > "More Ratios..."
Background information
Refer to the following sections:
●
Chapter 4.3.5.1, "Wave quantities"
●
Meas softtool