DISTRIBUTED BY:
00702-39
OPERATING INSTRUCTIONS
OAKTON WD-35606-00, -05
CON 5 and TDS 5
Acorn Series Meters
Printed in the U.S.A. 8/98
CON 5Acorn series
Conductivity/
°
C Meter
ON
OFF
CAL
HOLDENTER
MODE
Страница 1: ...DISTRIBUTED BY 00702 39 OPERATING INSTRUCTIONS OAKTON WD 35606 00 05 CON 5 and TDS 5 Acorn Series Meters Printed in the U S A 8 98 CON 5 Acorn series Conductivity C Meter ON OFF CAL HOLD ENTER MODE...
Страница 2: ...he Meter on 7 3 4 Setting the Conductivity to TDS conversion factor Acorn TDS5 only 7 4 Calibration 8 10 4 1 Conductivity TDS Calibration 8 9 4 3 Temperature Calibration 10 5 Measurement 11 5 1 Taking...
Страница 3: ...des the Specifications Accessories Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators m S t HO LO C pp m indicates opera...
Страница 4: ...he socket The phono jack connec tion can be removed with a simple pull 7 6 3 1 A A BNC and phono jack connections at top of Acorn meter 3 2 Switching the meter on Press the ON OFF key to power up your...
Страница 5: ...y 2 3 the full scale F S value of the measurement range For example in the 200 to 1999 S range use a 1413 S solution for calibration Note the Acorns will not accept calibration values less than 40 S 2...
Страница 6: ...hows the correct temperature Maximum adjustment if 5 C from the factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter then reverts to measurement mode A A C B B 4...
Страница 7: ...gitated mild detergent bath Wipe probe with soft tissue paper Rinse thoroughly in tap water and then in deionized water Recalibrate the meter after cleaning the probe 7 Error Messages Error Message in...
Страница 8: ...between 40 S acceptable calibration range 20 ppm and 19 9 mS 9 9 ppt 8 Specifications Mode Conductivity TDS Temperature CON 5 TDS 5 0 0 to 199 9 S 0 0 to 99 9 ppm Range 200 to 1999 S 100 to 999 ppm 0...
Страница 9: ...S values 757 1 ppm KCl 737 1 ppm NaCl 1050 ppm 442 WD 00653 27 Conductivity 2070 S solution TDS values 1045 ppm KCl 1041 ppm NaCl 1500 ppm 442 WD 00653 20 Conductivity 2764 S solution TDS values 1382...
Страница 10: ...and correction will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 16 Ret...