4. Calibration

Conductivity/TDS calibration

This instrument is capable of up to three-point calibration to ensure accuracy across
the entire range of the meter. You need only one calibration for measurement of the
entire range of the meter. If a range is not calibrated, the meter automatically detects
the closest range calibrated and uses the calibration information. However, only 
the ranges that have been calibrated have maximum accuracy. 

Calibrate to all

ranges to ensure highest accuracy throughout all measurement ranges.

When to calibrate

If you are measuring in solutions with conductivity lower than 100 µS or TDS lower
than 50 ppm, calibrate the meter at least once a week to get specified ±2% full scale
accuracy. If you are measuring in the mid ranges and you wash the probe in deion-
ized water and store it dry, calibrate the meter once a month. If you take measure-
ments at  extreme temperatures, calibrate at least once a week

Use only the supplied conductivity/TDS probe or replacement probe WD-35606-50. 

NOTE

Do not reuse solutions after calibration. Contaminants in the solution can

affect the calibration, and the accuracy of the measurements.

Preparing for calibration

Select a conductivity or TDS standard near the sample value you are measuring. If
this is not possible, choose a calibration solution value that is approximately 2/3 
the full scale (F.S.) value of the measurement range. For example, in the 200 to 1999
µS range, use a 1413 µS solution for calibration.  Note: the Acorns will not accept 
calibration values less than 40 µS (20 ppm).

Wet the probe for 10 minutes prior to calibration to saturate the probe and 
minimize drift. 

It is not necessary to reset the meter before calibration. All new calibrations will
automatically override existing calibration. 

Calibrating for Conductivity or TDS

1.

Pour approximately 3 cm of the chosen Conductivity / TDS standard into 
two separate containers.

2. 

Gently stir the probe in one of the containers of your standard. This standard 
acts as a rinse solution, removing contami-
nants that offset the calibration.

3. 

Dip the rinsed probe into the other container
of your standard. 

4.

Turn meter on. Meter will automatically enter
Conductivity/TDS mode.

5. 

Tap probe on the bottom of the container to
remove air bubbles. Let the probe stabilize to
the solution temperature.

6. 

Press the CAL key. The meter will blink with
the factory calibrated value.

See figure

7. 

Press 

(up) or 

(down) key until the 

displayed value matches your standard 
solution.

8. 

Press enter key to confirm. The meter will exit 
calibration, and return to measurement mode.

Repeat for all ranges for maximum accuracy.

4.1

m

S

A

B

A

B

Range

Resolution

Accuracy

0.0 to 199.9 µS*

0.1 µS

±4 µS

Conductivity

200 to 1999 µS

1 µS

±40 µS

2.00 to 19.99 mS

0.01 mS

±0.4 mS

0.0 to 99.9 ppm*

0.1 ppm

±2 ppm

TDS

100 to 999 ppm

1 ppm

±20 ppm

1.00 to 9.99 ppt

0.01 ppt

±0.2 ppt

*Minimum calibration value 40 µS (20 ppm).

Содержание WD-35606-00, WD-35606-05

Страница 1: ...DISTRIBUTED BY 00702 39 OPERATING INSTRUCTIONS OAKTON WD 35606 00 05 CON 5 and TDS 5 Acorn Series Meters Printed in the U S A 8 98 CON 5 Acorn series Conductivity C Meter ON OFF CAL HOLD ENTER MODE...

Страница 2: ...he Meter on 7 3 4 Setting the Conductivity to TDS conversion factor Acorn TDS5 only 7 4 Calibration 8 10 4 1 Conductivity TDS Calibration 8 9 4 3 Temperature Calibration 10 5 Measurement 11 5 1 Taking...

Страница 3: ...des the Specifications Accessories Warranty and Return of Items section 2 Display and Keypad Functions The Acorn has a large custom LCD with the following indicators m S t HO LO C pp m indicates opera...

Страница 4: ...he socket The phono jack connec tion can be removed with a simple pull 7 6 3 1 A A BNC and phono jack connections at top of Acorn meter 3 2 Switching the meter on Press the ON OFF key to power up your...

Страница 5: ...y 2 3 the full scale F S value of the measurement range For example in the 200 to 1999 S range use a 1413 S solution for calibration Note the Acorns will not accept calibration values less than 40 S 2...

Страница 6: ...hows the correct temperature Maximum adjustment if 5 C from the factory default 6 Press ENTER key to confirm calibration The LCD displays CO and the meter then reverts to measurement mode A A C B B 4...

Страница 7: ...gitated mild detergent bath Wipe probe with soft tissue paper Rinse thoroughly in tap water and then in deionized water Recalibrate the meter after cleaning the probe 7 Error Messages Error Message in...

Страница 8: ...between 40 S acceptable calibration range 20 ppm and 19 9 mS 9 9 ppt 8 Specifications Mode Conductivity TDS Temperature CON 5 TDS 5 0 0 to 199 9 S 0 0 to 99 9 ppm Range 200 to 1999 S 100 to 999 ppm 0...

Страница 9: ...S values 757 1 ppm KCl 737 1 ppm NaCl 1050 ppm 442 WD 00653 27 Conductivity 2070 S solution TDS values 1045 ppm KCl 1041 ppm NaCl 1500 ppm 442 WD 00653 20 Conductivity 2764 S solution TDS values 1382...

Страница 10: ...and correction will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 16 Ret...