GUI reference
R&S
®
ZNA
831
User Manual 1178.6462.02 ─ 20
Initially a set of preconfigured converter types is available, mainly comprising the leg-
acy converter models R&S
ZVA-Zxxx(E). To define your own converter types, proceed
as follows:
1. Use "Add" to create a converter type with default properties or use "Copy" to dupli-
cate the properties of the converter type whose tab is currently selected.
2. Enter a (unique) name for the new converter type.
3. Modify the converter properties to suit the converter type you want to introduce to
the R&S
ZNA.
Preconfigured converter types are read-only.
Defined Converter Types (Tabs)
The tabs in the "Converter Types" represent the converter types that are currently
defined (pre-defined and user-defined).
Remote command:
[SENSe:]CONVerter:DEFinition:CATalog?
[SENSe:]CONVerter:DEFinition:COUNt?
Type
The name of the converter type.
Remote command:
[SENSe:]CONVerter:DEFinition:CATalog?
[SENSe:]CONVerter:DEFinition:DEFine
IF
Defines the fixed IF frequency of the converter type.
Remote command:
[SENSe:]CONVerter:DEFinition:IFRequency
Converter Port: Start Freq./Stop Freq.
Defines the start and stop frequency of the (converter port of the) converter type.
Remote command:
[SENSe:]CONVerter:DEFinition:FREQuency
Source: Multiplicator
Defines the source multiplicator, i.e. the factor by which the incoming RF frequency is
multiplied by the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:SMUL
Source: Nominal Power
Defines the preferred input power at the RF In port of the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:SPOWer