Command reference
R&S
®
ZNA
1298
User Manual 1178.6462.02 ─ 20
Example:
Suppose that the cal pool contains a file
'Calgroup3.cal'
with a valid through calibration for the active channel no. 1,
which you want to include in a new TOSM calibration for ports 1
and 2.
MMEM:CDIR DEF; CDIR 'Calibration\Data'
Go to the cal group directory.
:SENSe1:CORRection:COLLect:METHod:DEFine 'New
Cal', TOSM, 1, 2
Define a new TOSM calibration.
SENSe1:CORRection:COLLect:LOAD:SELected 'Cal
Group 3.cal', THROUGH, 1, 2
Load the through data from the cal pool file into the new calibra-
tion.
SENSe1:CORRection:COLLect:ACQuire:SELected
OPEN, 1
Proceed with the new calibration measurements.
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalName>, <CalType>,
<TestPort1>[, <TestPort2>[, <TestPort3orAdapterCalKit>[, <TestPorts>]...]]
Defines the calibration to be performed for channel
<Ch>
.
If multiple channels shall be calibrated in parallel, use
●
[SENSe:]CORRection:COLLect:CHANnels:ALL
●
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
before executing this command. For background information see
lel calibration of multiple channels"
Calibration data is acquired using
[SENSe<Ch>:]CORRection:COLLect[:
.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName>
Name of the calibration (string parameter). The name serves as
a reference to delete a particular set of system correction data
(
[SENSe<Ch>:]CORRection:COLLect:DELete
<CalType>
ARTosm | ARPTosm | REFL | RSHort | FOPort | FTRans |
RTRans | FRTRans | OPTPort | TOSM | UOSM | VUOSm |
TRL | TOM | TSM | TRM | TNA | PTOSm | PTOM | PTSM |
PTRM | PTRL | PTNA | PUOSm | VPUosm | CPUosm |
MTOSm | MUOSm | MOSM | MTRL
Calibration types; see list below.
SCPI command reference