GUI reference
R&S
®
ZNA
665
User Manual 1178.6462.02 ─ 20
Access:
●
Channel – [Offset
Embed] > "Offset" > "Fixture Compensation..."
●
Channel – [Offset
Embed] > "One Way Loss" > "Fixture Compensation..."
Background Information
Refer to
Chapter 4.6.1.5, "Fixture Compensation"
Ports
Selects the ports for whom fixture compensation data shall be acquired.
Auto Length / Auto Length and Loss
"Auto Length" or "Auto Length and Loss" implies that a global electrical length offset
and loss is determined in analogy to the general offset compensation (see
on page 202 and
Chapter 4.6.1.4, "Auto Length and Loss"
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
Direct Compensation
With "Direct Compensation", a frequency-dependent transmission factor is calculated;
see
"Auto Length and Loss vs. Direct Compensation"
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Prompt for Each Port
Determines how the R&S
ZNA performs the sweeps for a given termination type (Open
●
If unchecked, it performs the sweeps for Open/Short without interruption, implicitly
assuming that
all
ports are terminated accordingly
●
If checked, it interrupts the data acquisition process after each port, which allows
you to modify the test setup (e.g. terminate the next measured port).
Measurement Type
The "Open" , "Short", and "Open and Short" buttons bring up the "
" dialog that guides you through the actual fixture measurement. See
vs. Open and Short compensation"
Measure Fixture wizard
The "Measure Fixture" dialog guides you through the previously configured fixture
compensation measurements.
Offset Embed softtool