8.5
8 - 5
8.5 Memorizing Optical Temperature Control Equation
Cuvette heating under optical temperature control is intended to
memorize the intensity of the light emitted from the cuvette at
each temperature and to control heating so as to match the light
intensity.
This is executed independently for each type of cuvette and
calibration data is retrieved for each cuvette. No equation is
memorized for a cuvette type used for the first time. In such an
event, click [Confirm GA optical temperature control] in the
[Instrument control] menu for the execution.
In atomization under the optical temperature control, once the
equation is memorized, atomization temperature is always kept
at a constant level (emission intensity) despite a change in
cuvette resistance.
Since the relationship has been memorized before shipment
from the factory, it needs not to be memorized frequently by user.
Memorization is required once in a few months or half-yearly
because the emission intensity received by the detector may
differ due to change of the instrument with time. Memorization
of the equation involves heating for about 2 minutes.
In this case, be sure to use the Pylon tube HR, Pylon tube C HR
or Tube HR cuvette.
For memorization, click [Memorize Opt. Temp. Cont. Equation] in
the [Instrument Control] menu.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...