3.1.3
3 - 3
(2) In the periodic table (<1> in Fig. 3-2), click the elements to
be measured in sequence. The clicked element turns
green and is added to the [Element - Order] table (<2> in
Fig. 3-2). A maximum of 8 lamps can be set in the
instrument, and when using multi-element lamps, up to 12
elements can be set. Enter the lamp number shown on
the lamp socket.
(3) To delete an analyte element, use the following steps.
Click an element in the periodic table.
Click the element highlighted in yellow-green to change
the color to white.
To uncheck [Measurement] in the [Element-Order], click
the check mark
in the [Measurement] column.
The row will remain in the table, but measurement will
not be carried out.
Click
[Delete].
Click [No.] in the [Element-Order] column for the
element to delete to change the color of the row of the
element to blue (<3> in Fig. 3-2). Click the [Delete]
button.
Click the [Delete All] button to delete all elements in the
[Element-Order] column.
(4) Click a [Lamp No.] in the Element-Order table and enter the
turret number of the lamps set for each element.
(5) Measurement will be carried out in the numerical order of
[No.] entry. For changing the order of a specific element,
click the [No.] column of the element to turn its row blue.
Then, move the row by the [Up] or[ Down] button.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...