5.6
5 - 17
2) The computer monitors the percentage of the signal
convergent width. The signal convergent width is
defined as the change in signal for 1 second divided by
the corresponding mean absorbance. If the signal
convergent width is within the specified range, the
computer carries out measurement after waiting for the
specified delay time.
The signal convergent width is specifiable at four
levels according to the degree of absorbance, and the
delay time is specifiable at two levels.
3) At the end of the current measurement, the signal
becomes lower than the end threshold level.
Then, the computer recognizes that the current
measurement has been completed, and it waits for
injection of the next sample.
Measurement
Delay time
Signal convergent width
Start threshold level
End threshold level