5.1 Various Measurement Methods
5 - 8
(3) In [Determination Mode], select [Simple STD Add.], and in
[Order], select [Linear].
(4) In the [Working Curve] column, make entries for [No. of
STDs], [STD Replicate], [STD Decimal Place], [STD Unit]
and STD [Conc.].
In the standard addition method, concentration is calculated
by extrapolation of a linear working curve. Therefore, STD
concentration should be set within a range where working
curve is neither concave nor convex. In addition, STD1
concentration is required to be 0.
(5) Make entries for [UNK replicate], [UNK Decimal Place] and
[UNK Unit]. Note that [UNK Replicate] is applied to the
second sample onward. (As the first sample, a standard
addition sample is measured according to the STD
replicate count.)
(6) Make entries in the [Sample Table] window. In [No. of
UNK], set the number of samples including the first
unknown sample.
(7) After verifying the method specified in the above steps,
start measurement.
NOTE
: For the first unknown sample, select the one that
contains a matrix (coexisting substance) constituent
having the average composition of all unknown
samples and whose absorbance is not extremely high.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...