5.8 Correction of Working Curve Drift (Reslope function)
5 - 34
(5) Measurement of STD4
Creation of working curve by resloping
A
1
: Initial STD4 absorbance
A
3
: STD4 absorbance in reslope measurement
(6) Measurement of Unknown Samples 31 to 40 (using
working curve in (5))
S1 S2 S3 S4 S5 S6
A1
A2
A3
n
i
i
i
C
a
A
0
n
i
i
i
C
a
A
A
A
0
1
2
n
i
i
i
C
a
A
A
A
0
1
3
Setting is required as follows.
(a) Open the Standards Table window and turn on the
[Reslope] check box.
(b) In [STD for Reslope], enter the name of standard
sample used for correction.
(c) In [Reslope Interval], enter the sample interval for
correction.
Supplement: Resloping in manual measurement
Usually, reslope measurement is carried out at the set
sample interval. Also, this measurement can be
started freely by clicking the [Start] button in
succession to the [Reslope] button in the tool box.
Fig. 5-31 [Reslope] Button
n
i
i
i
C
a
A
A
A
0
1
3
A
bs
or
banc
e (
A
)
Concentration (C)
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...