3 - 6
(1) Detach the fitting 1 and 2 by loosening the fitting screws.
Fig. 3-1 Absorption Cell and Cell Holder
(2) Attach the absorption cell to the lower side of the cell holder and
fix it with the fittings. Be aware that upper/lower shape of fitting 1
is different.
At this time, do not tighten the fitting screws so as to adjust the
cell position.
Fig. 3-2 Assembled Absorption Cell and Silicone Tube
Fig. 3-3 Attachment of Fitting 1
Absorption cell
Fitting 2
Cell holder
Holder-fixing screw
Fitting screws
Fitting 1
Silicone tube