1 - 11
Fig. 1-11 Setting of Pump Tubing for Four-liquid Flow Channel
NOTE 1:
Each tube should be inserted into the nipple completely. If
connection is incomplete, poor suction or leakage may
occur.
NOTE 2:
Each tube should be connected correctly in accordance
with the identification marker.
(3) The nipple of each tube connected in the process of above (2)
has a groove on it. Set the nipple groove on the OUTLET side
tube guide in the order of the sodium borohydride suction tube,
hydrochloric acid suction tube, sample suction tube, and
reducing agent suction tube. Then, on the INLET side, set the
nipple groove on the tube guide while aligning each tube with
the groove on the pump rotor.
(4) Remove the head cap and inner cap from each of the acid bottle
(part No. 171-6336), sodium borohydride bottle (part
No.171-6337), reducing agent bottle (part No. 7J1-7338), and
connect them to the hydrochloric acid suction tube, sodium
borohydride suction tube, and reducing agent suction tube,
respectively. Put these tubes into the respective bottles, set the
caps and place the bottles on the reagent tray (part No.
171-6395), which is provided as an accessory.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...