4 - 6
NOTE 2:
Although the color of the thermal tape on the flame
magnet may be changed to black, that is not abnormal.
NOTE 3:
Do not extinguish the flame without disconnecting the
silicone tube from the separator. Otherwise water is
introduced into the absorption cell, and the cell may
crack.
CAUTION
The absorption cell and cell holder become very
hot when the flame is ignited. There is a danger of
burns upon touching them.
Do not touch the cell and cell holder when the
flame is ignited. Wait until the cell and cell holder
cool down before handling them.
(8) Attach the end of the silicone tube, the other end of which is
attached to the absorption cell, to the separator.
(9) Wait for about 10 minutes until the flow rate of the solutions
become constant.
4.5.2 Operation of Measurement Using Four Liquids
(1) Perform the measurement operation using the four-liquid flow
channel. Refer to “3.7 Measurement”.
(2) After measurement of the last sample, perform the closing
operation. Refer to “3.8 Closing Operations”.
(3) When using the autosampler, delay time of the the four-liquid
flow channel will be longer than the three-liquid flow channel by
the length of the reaction coil.
Refer to “3.9 Measurement with Autosampler”.
(4) Automatic addition of the pre-reducing agent by the four-liquid
flow channel is not recommended for elements other than
arsenic.
(5) In the arsenic analysis, the pre-reduction of some samples is
insufficient using the four-liquid flow channel.
If there is a problem with the measurement using the four liquid
flow channel, it should be measured using the three liquid flow
channel after at a sufficient reaction time with the addition of
pre-reducing agent during sample preparation.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...