9.3.6
9 - 13
(2) Click the [Print Preview] button, and the print preview will
be displayed.
Clicking the [Output] button will output the diagnostic data
to the set destination.
NOTES
: 1. The spectrum of each wavelength displayed for
the wavelength accuracy shows the status at the
time of diagnosis. It does not show the sensitivity
or differences. Check “Judgment results” in the
table for the diagnosis results.
2. The method of setting wavelength used for
instrument diagnosis moves directly to the
wavelength set. This method is dependent on the
accuracy of the wavelength drive mechanism. It is
not necessarily positioned at the peak of the lamp
energy.