3.5 Preparation for Autosampler
3 - 58
(5) Click the [Move Nozzle] button on the toolbar again for
inserting the nozzle tip into the sample injection port in
graphite cuvette.
(6) Click the [Move Nozzle] button on the toolbar again for
returning the nozzle to the rinse cup position.
NOTE
: If the cuvette placed is different from the one specified
in the Method window, click the [Reset Autosampler]
button
to return the nozzle to the previous position.
If [Move Nozzle] is clicked, the nozzle may not enter
the sample inlet and may be damaged.
Prepare standard samples and unknown samples, and set them
on the sample tray according to the sample map in the
Autosampler display of Method window.
Sample cups can be set in Nos. 1 to 60. The cups are numbered
1 to 10 from bottom left to right and numbered 11 to 20 on one
level up. The top right one is numbered 60.
Fig. 3-44 Sample Cups
Sample cup
: minimum volume
0.4 mL
maximum volume 1.5 mL
Reagent container : minimum volume
5 mL
maximum volume
17 mL
Sample cup
Reagent container
3.5.4 Setting of Sample
Cups
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...